A Comparison of the Image Quality and Radiation Dose Using 100-kVp Combination of Different Noise Index and 120-kVp in Computed Tomography Pulmonary Angiography
Objective The objective of our study was to compare the image quality and radiation dose of computed tomography pulmonary angiography (CTPA) in patients with different body mass indexes using 100-kVp combination of different noise indexes (NIs) and 120-kVp scan protocol with the adaptive statistical iterative reconstruction 2.0 algorithm (ASiR 2.0). Methods A total of 120 patients who had undergone a CTPA were divided into 4 groups (A, B, C, and D), with 30 patients in each group. Group A underwent 120-kVp CT scan protocol in combination with NI = 25, while groups B, C, and D underwent 100-kVp CT scan protocol in combination with NI = 30, 35, and 40, respectively. All images were restructured using ASiR 2.0. The signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) of all images were calculated when the CTPA was completed. Each subjective image evaluation used a 5-point scoring method and was conducted by 2 independent radiologists. The CT dose index of volume and dose-length product were recorded, and the mean value was calculated. The dose-length product was converted to the effective dose. Results There were no significant differences in SNR, CNR, and subjective image quality among the groups A, B, C, and D. The effective dose of group D decreased by 48.33% and 27.27% relative to groups A and B, respectively (P < 0.01). Conclusions The 100-kVp CT scan protocol in combination with NI = 40 can more effectively reduce the radiation dose than can the 120-kVp CT scan protocol in combination with NI = 25 for a CTPA while still maintaining diagnostic image quality.
基金:
National Training Program of Innovation and Entrepreneurship for Undergraduates of China [201510075026]; Hebei Province Program of Training and Basic Project of Clinical Medicine of China; Affiliated Hospital of Hebei University Outstanding Youth Foundation [2015Q002, 2015Q017]
第一作者机构:[1]Tianjin Univ, Sch Life Sci, Sch Precis Instrument & Optoelect Engn, Tianjin, Peoples R China[2]Hebei Univ, Affiliated Hosp, Dept Radiol, Baoding, Peoples R China
通讯作者:
通讯机构:[4]Tianjin Univ, Sch Life Sci, 92 Weijin Rd, Tianjin 300072, Peoples R China[*1]School of Life Science, TianjinUniversity, 92 Weijin Rd, Nankai District, Tianjin, 300072 People'sRepublic of China
推荐引用方式(GB/T 7714):
Zhao Yong-Xia,Zuo Zi-Wei,Suo Hong-Na,et al.A Comparison of the Image Quality and Radiation Dose Using 100-kVp Combination of Different Noise Index and 120-kVp in Computed Tomography Pulmonary Angiography[J].JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY.2016,40(5):784-790.doi:10.1097/RCT.0000000000000488.
APA:
Zhao, Yong-Xia,Zuo, Zi-Wei,Suo, Hong-Na,Wang, Jia-Ning&Chang, Jin.(2016).A Comparison of the Image Quality and Radiation Dose Using 100-kVp Combination of Different Noise Index and 120-kVp in Computed Tomography Pulmonary Angiography.JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY,40,(5)
MLA:
Zhao, Yong-Xia,et al."A Comparison of the Image Quality and Radiation Dose Using 100-kVp Combination of Different Noise Index and 120-kVp in Computed Tomography Pulmonary Angiography".JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY 40..5(2016):784-790