A Comparison of the Image Quality and Radiation Dose Using 100-kVp Combination of Different Noise Index and 120-kVp in Computed Tomography Pulmonary Angiography

文献详情

资源类型:
WOS体系:
Pubmed体系:

收录情况: ◇ SCIE

机构: [1]Tianjin Univ, Sch Life Sci, Sch Precis Instrument & Optoelect Engn, Tianjin, Peoples R China [2]Hebei Univ, Affiliated Hosp, Dept Radiol, Baoding, Peoples R China [3]Hebei Univ, Sch Med, Baoding, Peoples R China [4]Tianjin Univ, Sch Life Sci, 92 Weijin Rd, Tianjin 300072, Peoples R China
出处:
ISSN:

关键词: low tube voltage noise index radiation dose image quality computed tomography pulmonary angiography

摘要:
Objective The objective of our study was to compare the image quality and radiation dose of computed tomography pulmonary angiography (CTPA) in patients with different body mass indexes using 100-kVp combination of different noise indexes (NIs) and 120-kVp scan protocol with the adaptive statistical iterative reconstruction 2.0 algorithm (ASiR 2.0). Methods A total of 120 patients who had undergone a CTPA were divided into 4 groups (A, B, C, and D), with 30 patients in each group. Group A underwent 120-kVp CT scan protocol in combination with NI = 25, while groups B, C, and D underwent 100-kVp CT scan protocol in combination with NI = 30, 35, and 40, respectively. All images were restructured using ASiR 2.0. The signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) of all images were calculated when the CTPA was completed. Each subjective image evaluation used a 5-point scoring method and was conducted by 2 independent radiologists. The CT dose index of volume and dose-length product were recorded, and the mean value was calculated. The dose-length product was converted to the effective dose. Results There were no significant differences in SNR, CNR, and subjective image quality among the groups A, B, C, and D. The effective dose of group D decreased by 48.33% and 27.27% relative to groups A and B, respectively (P < 0.01). Conclusions The 100-kVp CT scan protocol in combination with NI = 40 can more effectively reduce the radiation dose than can the 120-kVp CT scan protocol in combination with NI = 25 for a CTPA while still maintaining diagnostic image quality.

基金:
语种:
被引次数:
WOS:
PubmedID:
中科院(CAS)分区:
出版当年[2017]版:
大类 | 4 区 医学
小类 | 4 区 核医学
最新[2025]版:
大类 | 4 区 医学
小类 | 4 区 核医学
JCR分区:
出版当年[2016]版:
Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
最新[2023]版:
Q4 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING

影响因子: 最新[2023版] 最新五年平均 出版当年[2016版] 出版当年五年平均 出版前一年[2015版] 出版后一年[2017版]

第一作者:
第一作者机构: [1]Tianjin Univ, Sch Life Sci, Sch Precis Instrument & Optoelect Engn, Tianjin, Peoples R China [2]Hebei Univ, Affiliated Hosp, Dept Radiol, Baoding, Peoples R China
通讯作者:
通讯机构: [4]Tianjin Univ, Sch Life Sci, 92 Weijin Rd, Tianjin 300072, Peoples R China [*1]School of Life Science, TianjinUniversity, 92 Weijin Rd, Nankai District, Tianjin, 300072 People'sRepublic of China
推荐引用方式(GB/T 7714):
APA:
MLA:

资源点击量:15100 今日访问量:0 总访问量:960 更新日期:2025-05-01 建议使用谷歌、火狐浏览器 常见问题

版权所有©2020 河北大学附属医院 技术支持:重庆聚合科技有限公司 地址:保定市莲池区裕华东路212号