Improvement of Abdominal Imaging Quality by Optimal Adaptive Statistical Iterative Reconstruction Combined with Monochromatic Spectral Computed Tomography Imaging

文献详情

资源类型:
WOS体系:

收录情况: ◇ SCIE

机构: [1]Hebei Univ, Affiliated Hosp, CTIMRI Room, Baoding 071000, Hebei, Peoples R China [2]Heibei Med Univ, Hosp 2, Dept Med Imaging, Shijiazhuang 050000, Hebei, Peoples R China [3]Shijiazhuang First Hosp, Shijiazhuang 050011, Hebei, Peoples R China
出处:
ISSN:

关键词: Spectral Computerized Tomography Angiography Monochromatic Image Adaptive Statistical Iterative Reconstruction Contrast to Noise Ratio

摘要:
Purpose: To investigate the optimal adaptive statistical iterative reconstruction (ASIR) percentage in monochromatic level combined with spectral computed tomography imaging for improving abdominal image quality. Methods: The monochromatic images at 55 and 70 keV in 32 patients with abdominal diseases at arterial and venous phases, noise, contrast-to-noise ratio (CNR) and subjective scores of image quality were analyzed. The reconstruction imaging was performed at 0%, 30%, 50% and 70% ASIR for the arterial and portal venous phases in the tested group. The data of these images were analyzed after a total of le sets of data were reconstructed at 70 keV and 55 keV energy points with different ASIR percentages. Results: In 50 and 70 keV, the noise was significantly decreased with increase of ASIR percentage. In 70 keV, the liver CNR in the arterial phase and the pancreatic CNR in both phases were significantly greater in 70% than in all the other percentage ASIR. The artifacts were significantly fewer in the 30%, 50% and 70% ASIR than in the 0% ASIR, and also fewer in 30% ASIR than in 70% ASIR in both phases. The diagnostic confidence was significantly greater in 30% and 50% ASIR than in 0% and 70% ASIR in the arterial phase and was significantly greater in 30%, 50% and 70% ASIR than in 0% ASIR in the venous phase. In the 55 keV, the left kidney CNR was significantly greater in 50% or 70% ASIR than in 0% or 30% ASIR, and the venous phase pancreatic CNR was significantly greater in 70% ASIR than in 0% or 30% ASIR. The vessel images were significantly better in the 50% than in the all other percentage ASIR. Conclusion: This study shows that 70 keV plus 30% ASIR and 55 keV plus 50% ASIR are the optimal protocol which can improve the image quality of the abdominal organs and blood vessels.

基金:
语种:
被引次数:
WOS:
中科院(CAS)分区:
出版当年[2020]版:
大类 | 4 区 医学
小类 | 4 区 数学与计算生物学 4 区 核医学
最新[2025]版:
JCR分区:
出版当年[2019]版:
Q4 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Q4 MATHEMATICAL & COMPUTATIONAL BIOLOGY
最新[2023]版:

影响因子: 最新[2023版] 最新五年平均 出版当年[2019版] 出版当年五年平均 出版前一年[2018版]

第一作者:
第一作者机构: [1]Hebei Univ, Affiliated Hosp, CTIMRI Room, Baoding 071000, Hebei, Peoples R China
共同第一作者:
通讯作者:
推荐引用方式(GB/T 7714):
APA:
MLA:

资源点击量:15101 今日访问量:2 总访问量:962 更新日期:2025-05-01 建议使用谷歌、火狐浏览器 常见问题

版权所有©2020 河北大学附属医院 技术支持:重庆聚合科技有限公司 地址:保定市莲池区裕华东路212号